HIGH PRESSURE FOAM ANALYZER HPFA

The HPFA High Pressure Foam Analyzer is the world’s only measuring instrument for simultaneously analyzing the amount and structure of liquid foams under high pressure. The instrument provides various options for investigating foam behavior under the real process conditions of foam-assisted flooding methods in EOR as well as hydraulic fracturing.

Additional information
INSTRUMENT DIMENSIONS

Footprint: 1100 mm × 640 mm (W × D)
Height: 1070 mm
Weight (without accessories): 190 kg

ENVIRONMENT

Operating temperature 15 to 30 °C
Humidity without condensation

TEMPERATURE MEASUREMENT

Sensor: PT100
Range of temperature measurement: 4 to 90 °C
Resolution: 0.1 °C
Precision: 0.1 °C
Accuracy: 1/3 DIN B (±0.1 °C at 0 °C / ±0.8 °C at 400 °C)
Location of temperature measurement: inside sample liquid
External sensor: optional

TEMPERATURE CONTROL

Type of temperature control: optional: double-walled glass column
Range of temperature control: 4 to 90 °C
(additional thermostat needed: TB14)

ACCESSORIES

ilter plates for sparging: Diameter: 30 mm
Filter plate porosities: G2 – nominal maximum pore size – 40 to 100 μm (recommended)
G3 – nominal maximum pore size – 16 to 40 μm
G4 – nominal maximum pore size – 10 to 16 μm
Material of columns and frits: borosilicate glass (norm: ISO 4793)
Material of sealings: PTFE with glass fibers (inner) and graphite (outer) for the windows
HNBR for the filter plates

TEMPERATURE SENSOR

Sensor: 1 thermocouple
Location: in the high-pressure cell (top/middle/bottom)

MINI DOSING SYSTEM

Dosing: manual
Location: in the high-pressure cell (top/middle/bottom)
Temperature: maximum 180 °C

PRESSURE MEASUREMENT

Maximum pressure: 350 bar (5000 psi)
(additional external syringe pump needed)

ILLUMINATION

Type: LED
Wave length, dominant (height detection): 469 nm
Wave length, dominant (structure detection): 633 nm

STRUCTURE CAMERA

Connection: USB 2.0
Performance: 2 fps at 1280 × 1024 px
Mean field of view: 25 × 13 mm²
Diameter of minimum detectable bubble: 150 μm
Focus: manual

HEIGHT CAMERA

Connection: USB 2.0
Performance: 5 fps at 1280 × 1024 px
Spatial resolution: 200 dpi, 0.12 mm/px
Mean field of view: 85 × 5.2 mm²
Focus: manual

MEASUREMENT SPECIFICATIONS

foam capacity (additional flow sensor needed)
maximum foam density
expansion rate
foam half life time
drainage half life time
sample temperature
gas temperature
pressure
mean bubble area
bubble

Download PDF