Measuring range |
0 to 100 vol% Gas |
---|---|
Accuracy |
approx. ±0.1 vol% Gas |
Measuring media & Flow rates |
electrically non-conductive fluids |
Measuring time |
Standard: approx. 1 s, |
Temperature measurement |
PT 100, ± 0.1 °C |
Temperature range |
−30 °C to 150 °C |
Pressure range |
max. 10 bar |
Dimensions – measuring tube (WxHxL) |
Standard sensor: 40 x 110 x 165 mm; |
Material – measuring tube |
special steel (1.4571) |
Temperature range sensor |
-20°C to 120°C (optionally up to 300°C) |
Hydraulic connections |
Standard sensor: 2x UNF 9/16"-18Gg (dash-6) male |
Dimensions – electronic unit (WxHxD) |
Standard (Aluminum): 235 x 150 x 370 mm |
Data communication (computer) |
RS-232 interface (CGS software for Windows included) |
Analog output – temperature |
4 … 20 mA/0 … 10 V |
Analog output – CGp-signal |
4 … 20 mA/0 … 10 V |
Analog output – CG0-signal |
4 … 20 mA/0 … 10 V |
Analog output – pressure |
4 … 20 mA/0 … 10 V |
Power supply |
110 – 240 VAC or 9 – 36 VDC |
CGS Inline Aeration Meter
The CGS (Concentration Gas System) is an inline measuring system for the continuous determination of the gas content in fluids. The system works regardless of the size and shape of the entrained gas portions in the
fluid. Therefore both finely dispersed microbubbles as well as large air pockets can be detected correctly in the passing medium, with their percentage by volume displayed in real time.
The following parameters are determined by the CGS:
• temperature (−30 °C to 150 °C)
• pressure (max. 10 bar)
• gas content CGp at measured pressure
• gas content CG0 at atmospheric pressure (recalculation according to the ideal gas law)